30 de septiembre de 2006
Original Text: Bipolar Resectoscope

"Bipolar Resectoscope: The future perspective of hysteroscopic."

Authors: Luca Mencaglia *, Emmanuel Lugo, Cristiana Barbosa
* Florence Center of Ambulatory Surgery, Florence, Italy.

Introduction

The gynecological resectoscope, born from its urological equivalent, is commonly used in gynecological practice to resects or remove intracavitary pathology and also to perform endometrial ablation. The resectoscope consist of a telescopes (2,9 and 4 mm, preferably with a 12° angle of view to always keep the loop within the viewing field), an electrical loop to perform passive cuts and 2 sheaths for continuous flow suction and irrigation of liquid distension medium. Besides the cutting loop, other instruments such as micro-knives or series of coagulation or vaporization electrodes of various shapes can be connected to the resectoscope.

There are essentially two types of resectoscopes which differ in outer diameter: 7,5 mm and 9,2 mm. The 7.5 mm resectoscope should be used in case of narrow cervical canal or difficult dilatation, the 9,2 mm resectoscope allows to perform major surgery. 1

Electro surgery in hysteroscopy

Biological tissue contains a more or less high concentration of electrolytes, making it sufficiently conductive to be treated by electro surgically. The thermal effect of high frequency current is used for separating (cutting) and coagulating tissue (desiccation of tissue). The high frequency currents must be used on the human since low-frequency currents can stimulate nerve and muscle cells in the currents flow due to electrochemical processes (electrolysis). These effects are small enough to be disregarded with frequencies above 100 kHz. 2

Monopolar or bipolar current system can be adopted:

Monopolar Resectoscope:
Conventional hysteroscopic surgery uses a monopolar electrocautery system in which the current passes from the active electrode through the patient's body towards the return plate. The distension media used is Glycine 1.5% o sorbitol-mannitol (non electrolyte irrigation fluid). The monopolar resectoscope is connected to monopolar electrosurgery generator of high frequency, automatically controlled by an acoustic alarm system. In a monopolar system, the electrons flows from a electrosurgery generator to active electrode (electrode of the loop). From electrode, the current flow is transmitted to tissue, after to the plate (neutral electrode) and return to generator. This system is potentially dangerous since the electrons flows through the body, outside the surgeon visual control, before it can return to the generator. The new generators, however, decreased the incidence of electric damage. In these generators the cut current flow is automatic regulated based on tissue resistance. The unipolar loop can be used as coagulation, cut and combined (coag-cut) current. The coagulation current/flow is characterized of a intermittent current flow periods, which cause cellular dehydratation, resulting in tissue hemostasis. The non modulated cut flow is a continuous flow, with high intracellular temperature, causing cellular explosion. Non-modulated flow can be used also for coagulation and it should be preferred because the voltage is lower and continuous. 1

Bipolar Resectoscope:
In bipolar electro surgery the current flow through the tissue is restricted to the area between the two electrode's loops that are under visual control of the surgeon. In this case as distension media can be used saline solution, because no risk of current dispersion. The generator produces a high initial voltage spike that establishes a voltage gradient in a gap between the bipolar electrodes. When the activated bipolar electrode is not in contact with the tissue, the electrolyte solution in the uterus dissipates it. When the loop is sufficiently close to tissue, the high bipolar voltage spike arc between the electrodes converts the conductive sodium chloride solution into a non-equilibrium vapor layer or “ plasma effect ” containing energy-charged sodium particles. Once formed, this plasma effect can be maintained at lower voltages (100-350 RMSV). 3 With tissue contact, there is disintegration of tissue via molecular dissociation. Energetic species of the charged ions from the plasma effect result in disruption of carbon-carbon and carbon-nitrogen bonds. There is also electron impact dissociation of water molecules into exited fragments of H+ and OH – ions. Bottom line is rupture of cell membranes which translates in to visible cutting. Clinically, there is a precise tissue effect with minimal collateral damage, as the charged ions have an estimated penetration depth in tissue of only 50 to 100 m-micron (0.5 - 1 mm) 4,5 . The depth of coagulation is determined principally by the electrode configuration and by the system design, as well as by the technique used by the operator (time and pressure of contact). 6

Equipment

The new working element developed by Karl Storz System, the Autocon II 400 (Figure 1), is for use in laparoscopy, resectoscopy and open surgery. It can be equipped with four different HF output sockets: monopolar, bipolar, multifunction and neutral electrode sockets. The user can assign a socket with the following parameters: cutting mode, coagulation mode or cutting and coagulation activation type. A mode is characterized by different generator setting such as frequency, peak voltage, modulation etc. A total of 9 monopolar and 8 bipolar modes are available. However, each specific socket is only available for a part of these modes. The user can also assign the following parameters to a mode: power limitation cutting, cutting effect, power limitation coagulation or coagulation effect.

Using bipolar system, in the Saline Time-C-Cut mode as the Saline Coagulation mode, the HF voltage form is unmodulated sinusoidal, the rated frequency is 350 kHz (at RL=500ohms) +/- 10%, the crest factor is 1.4, the rated load resistance is 50 ohms, the maxim HF peak voltage is 190 Vp, the number of power/effects are 8, the maxim power output 370 W+/- 20% and the parameter standard of use is effect setting 4 for cut and also effect setting 4 for coagulation. The average applied HF-power is approximately 130 Watts for resection. Its is important to point out that, this peak voltage are much lower compared with monopolar systems.


Figure 1 . Autocom II 400

The bipolar components (Figure 2,3) are compatible of existing resectoscope (optics, sheath). The sheat designed by Karl Storz- Endoskope is electrically completely isolated, the current return direct via the return electrode, prevent a current flow via the sheath, guarantying a high level of safety. Karl Storz has available a variety of cutting loops (GP, GPV, GD, GDV) in 24 Fr., but for the market there are still other bipolar loops for resection in gynecology and urology under development.

Figure 2 (left). The bipolar element of the resectoscope.

Figure 3 (right). The direct current return via the electrode prevent a current flow via the sheath.

Clinical Data

In 2000, Loffer FD 7 , reported a preliminary experience with the Versa Point bipolar resectoscope in gynecology, using in 15 patients, vaporizing electrode in saline solution distending medium, showing be effective in the removal of sub mucous myomas. Golan et al 8 in 2001 reviewed outcomes of operative hysteroscopy, using bipolar electrical energy (Versa Point) in saline solution in 116 women with intrauterine pathology and they proposed this new technique to potential replacement of the monopolar resection. Furthermore many studies had demonstrated their benefits in urology. Recently Singh H et al 9 reported the utility of bipolar system in the transurethral prostate resection (TURP) with less postoperative dysuria. Wendt-Nordahl 10 et al underlined in 2004 the advantage of the bipolar system in TURP, with the Vista System, and stressed the theoretical benefit to avoid the risk of TUR Syndrome.

Our preliminary experience in gynecology consist in 57 patients (Table I) treated by Karl Storz 26 Fr bipolar resectoscope and Autocon II 400 high frequency unit with parameters standard selected 180 (in effect 4) for cut and 120 (in effect 4) for coagulation. The resection loops are completely insulated from the sheath of the resectoscope and are reusable loops. Saline solution (NaCl 0,9%) was used as distension medium with no complications.

Indication

> 4cm

Myomas G0

9

1

Myomas G1

15

7

Myiomas G2

10

2

Endometrial Ablation

12

Polyps

8

2

Uterine Septum

3

2

Total

57

cm = centimeters

Table 1. Bipolar Resection in 57 patients

Cutting power and coagulation appears sensible better in comparison with monopolar resection. Especially the first cut in case of fibroid tissue doesn't give any problem, thanks to the plasma effect. Furthermore the vision during resection is not disturbed from the presence of air bubbles. We included in our series complex cases to evaluate the technical characteristics of the instruments, including an important number of uterine fibroids, that in some cases where more than 4 cm of size and G1 and G2 (partially or totally intramural localization). Results in terms of time of surgery, intra operatory bleeding, complete removal of the pathology were better if compared with traditional monopolar resection. (Fig. 4,5,6,7)


Figure 4. Endometrial polyp resection with bipolar resectoscope.


Figure 5. Metroplasty of uterine septum with bipolar resectoscope .


Figure 6. Endometrial ablation with bipolar resectoscope.


Figure 7. Bipolar roller ball.

Conclusion

In our experience and also in the analysis of the published data appears very clear that the bipolar resectoscope presents some advantages in comparison with the monopolar.

The current flow through tissue is restricted to the area between the two electrode's loops that are under direct vision of the surgeon. Current can be regulated at all times and set to the lowest possible: optimal flow of current for minimally invasive treatment. The plasma effect of bipolar current allows better cut and coagulation. In the monopolar technique the current passes through many tissues outside the surgeons' visual control before it can return to the generator. 5,11 The risk of thermal injuries at distant organs or tissues, by direct contact of instruments, imperfection of insulation or diffusion of the electric current, is reduced in the bipolar technique. 12,13 It has minor risk of interference on other electronic equipments (ECG, pace makers and others) simultaneously connected to the patient. 5,14 Furthermore there is reduced stimulation of peripheral nerves including the obturator nerve, because no flow of current through the body of the patient.

Distension media:
According to Kolmer et al 15 and Koshiba et al 16 incidence of overflow syndrome in gynecology and TUR syndrome in urology vary considerably in the literature, ranging from 0.18% to 10.9%. Mebust et al 17 reported in 3885 patients an incidence of 2% of TUR syndrome. Kudela et al 18 in 1996, reported the risk of fluid overload syndrome during hysteroscopy monopolar procedure and underline the necessity to adhere to safety measures which includes selection of a suitable medium (hypotonic electrolyte-free solutions -- Glycine or sorbitol-mannitol solution), control duration of surgery, respect correct surgical indications and procedure and specially perform continuous control of balance of the distension medium. Estes et al 19 in 2003 stand out the danger of hypotonic, electrolyte-free distension media and their potentiality to be absorbed in volumes large enough to cause hiponatremia and hipervolemia.

Main concerns in urological and gynecological conventional monopolar resection are fluid absorption with hiponatremia, hipervolemia and glycine toxicity. This syndrome is very dangerous for the patient leading to neurotoxic coma and death in the worst cases. Most of the morbidities of the overflow syndrome are related to the use of hypotonic non electrolyte irrigation fluid. For this reason require close and continuous peri operative monitoring of the balance of distension medium by a nurse and frequent laboratory investigations. Bipolar resection system, permits resection using saline solution. The use of saline solution for distension media of the uterine cavity, is the principal advantage of this technology so avoid use of hypotonic non electrolyte solution that can cause fluid overload during the surgical procedure. Saline solutions is easily metabolized, is not toxic and can be used with higher quantity and it is also less expensive than conventional hypotonic non electrolyte solutions. Singh et al 9 in a bipolar versus monopolar TURP randomized controlled study reported a significant difference in serum sodium concentration post operatively. In bipolar TURP the change in serum sodium was -1.2 mEq/L (not different from preoperative serum Na concentration), whereas in the monopolar group the mean decrease was 4.6 mEq/L. In three patients serum Na was >125 mEq/L at risk of TUR syndrome. However also balance of distension medium using saline solution should be under control. Starkman and Santucci 20 retrospectively reviewed 43 undergoing TURP, 18 consecutive patients treated with the monopolar TURP and 25 with bipolar TURPs. These investigators found an unexpected case of hiponatremia and pulmonary edema in a bipolar TURP patient. Patel et al. also expressed concern for potential problems with hipervolemia and hiponatremia. They suggest warming the saline solution and emptying the bladder from time to time during surgery. In our experience no overflow syndrome occurs.

Tissue alterations:
In traditional monopolar resection the tissue's electrical resistance creates temperature as high as 400°C which leads to desiccation with significant collateral and penetrative tissue damage 6 . High frequency current generated by a bipolar instrument tends to remain superficial; Luciano et al 4,5 reported a 0.5 - 1 mm depth of penetration compared with the 3-5 mm seen in monopolar system allowing a better control of the cut and lower possibility of accidental injury. The technique allows to maintain the current between the active electrode and the adjacent return electrode. Plasma effect of the loop prevents sticking effect onto it. In this case tissue damage is minimized and tissue temperature range from 40°C to 70°C. Improved tissue analysis secondary to reduce carbonization of tissue has been also reported with better histological interpretation. Improved visibility aiding the identification of surgical landmarks during procedure have been repeatedly reported.

Less bleeding during resection.
Optimized resection current provided by the Karl Storz Autocon II 400 allows a better coagulation during resection with reduced bleeding. Furthermore the coagulation capacity by him self is extremely more powerful in the bipolar system in comparison with monopolar. This avoid time consuming re-coagulation after resection for coagulation and contribute to close the superficial capillary vascularization reducing also intravasation.

Better visibility.
Minor air bubbles and less bleeding during resection allow a better vision during surgery reducing length of surgery and improving results.

Reduced costs:
Compatible components of existing monopolar Karl Storz resectoscopes (optics, sheaths). Resection loops are reusable and have duration and a cost comparable with the traditional monopolar one's.

The outcomes of studies in gynecology and urology with bipolar system demonstrate its versatility and the possibility of rapid replacement of "old" monopolar system. The bipolar system is technically superior, cost-effectiveness and safer in comparison with the monopolar system. If we take in consideration also the medical-legal aspects, it will be very dangerous to maintain the old system, especially in case of complications. Preliminary findings of decreased morbidity using bipolar system using saline solution, force us to consider others factors involved in possible complications as duration of surgery as well as the experience of the surgeon. These variables must be included in our future investigations. However the decrease of "theorical risk" of overflow syndrome that give the bipolar system doesn't allows to avoid close peri operative monitoring of distension medium balance and laboratory investigations.

Larger prospective randomized clinical trails examining cost-effectiveness and long-term outcome need to be performed, although it seems already very clear that this technology will replace in a near future the conventional monopolar electro surgery.

 

References

  1. Mencaglia L, Cavalcanti L. Histeroscopia Cirùrgica . Rio de Janeiro: UnionTask Press, 2004: 15-28
  2. Soderstrom RM. Principles of Eletrosurgery During Endoscopy. In:Sammarco MJ, Stovall TG, Steege JF. Gynecologic Endoscopy . Baltimore: Williams & Wilkins press, 1996:179-92
  3. Stadler KR, Woloszko J, Brown IG. Repetitive plasma discharges in saline solutions. Appl Phys Lett 2001; 79: 4503-4505
  4. Luciano AA, Whitman G, Maier DB, et al . A comparison of thermal injury, healing patterns, and postoperative adhesion formation following CO 2 laser and electro microsurgery. Fert. Steril 1987; 48:1025-9
  5. Luciano AA. Power sources. Obstet. Gynecol Clin. N. Am 1995; 22:423-43
  6. Patel A, Adshead J. Firts clinical experience with new transurethral bipolar prostate electrosurgery resection system: controlled tissue ablation (Coblation Technology). J Endour 2004; 18: 959-964
  7. Loffer, FD. Preliminary experience with the Versa Point bipolar resectoscope using a vaporizing electrode in a saline distending medium. J Am Ass Gyn Lapar 2004; 7:498-502
  8. Golan A, Sagiv R, Berar M, et al . Bipolar electrical energy in physiologic solution: a revolution in operative hysteroscopy. J Am Ass Gyn Lapar 2001; 8: 252-258
  9. Singh H, Desai M, Shrivastav P, et al . Bipolar versus Monopolar transurethral resection of prostate: randomized controlled study. J Endour 2005; Vol 19: 333-338
  10. Wendt-Nordahl G; Hacker A; Reich O, et al . The Vista System: a new bipolar resection device for endourological procedures: comparison with conventional resectoscope ; Eur Urol 2004; 46:86-90
  11. Riedel HH, Semm K. There is no place in gynecological endoscope for unipolar of bipolar high frequency current. Endoscopy 1982; 14:51-4
  12. Levy BS, Soderstrom RM, Dail DH, et al . Bowel injuries during laparoscopy. Gross anatomy and histology. J Reprod Med 1985; 30:168-72
  13. Di Giovanni M, Vasilenko P, Belsky D, et al . Laparoscopic tubal sterilization. The potencial of thermal bowel injuries. J Reprod Med 1990. 35:951-4
  14. Odell RC. Electro surgery: Principles and safety issues. Clin Osbtet Gyn 1995; 38:610-21
  15. Kolmer T, Norlen H. Transurethral resection of the prostate: a review of 1111 cases. Int Urol Neph 1989; 21:47-55
  16. Koshiba K, Egawa S, Ohori M, et al . T- Does transurethral resection of prostate pose a risk to life?22 years outcome. J Urol 1995; 153: 1506-1509
  17. Mebust WK, Holtgrewe HL, Cockett ATK, et al . Transurethral prostatectomy: Inmediate and postoperative complications: A cooperative study of 13 participating institutions evaluating 3885 patients. J Urol 1989; 141: 243-247
  18. Kudela M, Lubusky D, Dzvincuk P. Risk of fluid overload syndrome during hysteroscopy procedures; Cesk Gyn 1996; 61:291-3
  19. Estes CM, Maye JP. Severe intraoperative hyponatremia in a patient scheduled for elective hysteroscopy: a case report. AANA J 2003; 71: 203-5
  20. Starkman JS, Santucci R. Comparision of bipolar transurethral resection of the prostate with standart transurethral prostatectomy: shorter stay, earlier catheter removal and fewer complications. BJU Int 2005; 95: 69-71

 

Back to Newsletter